How do I replace the glass component of Radnoti Water-jacketed Flex Tubing?

On occassions it may be necessary to replacethe glass component of Radnoti's Water-jacketed Flex Tubing. To do this, follow the simple procedure outlined below.

Removing Glass component:
servlet.ImageServer?id=015w00000026Kqm&oid=00D20000000HncM

1.Unscrew120730(或120729-01and120731) connector and set aside
2.Unscrew160169black cap and set aside
3.Remove and discard orange ferrule130155-002. It may be necessary to cut the inner tubing down to achieve this.
4.Remove120727and120733white connector and set aside
5.Remove the Glass component130155-003and discard if broken. The glass and external tubing can be separated by placing in warm water if required.
6. If you cut the inner tubing in step 3, cut the outer tubing accordingly.

Replacing Glass component:


1.Fit new Glass piece to the external tubing (again heat up tubing to ease connection on glass piece if required) making sure internal tubing130155-001Xis through the middle and a sufficient length.
2.饲料内部油管(130155-001X) through white connector (120727/120733-002)
3.Secure new orange ferrule to the end of tubing (130155-001x) and then position into white connector (120727/120733-002) For a good fit secure with plumbers tape
servlet.ImageServer?id=015w00000026Kqw&oid=00D20000000HncM servlet.ImageServer?id=015w00000026KrQ&oid=00D20000000HncM





4.Reconnect Black cap (160169)
servlet.ImageServer?id=015w00000026KrV&oid=00D20000000HncM


5.Reconnect120730connector for completion.
servlet.ImageServer?id=015w00000026Kra&oid=00D20000000HncM